Model |
Probe Model 0203 (for middle temp.)
32 – 382F (0 – 200ºC) |
Probe Model 0204 (for high temp.)
32 – 752F (0 – 400ºC) |
|
Measuring Range |
40
– 9840fpm (0.2 – 50m/s): 32 – 212ºF
(0 – 100ºC) |
|
80
–9840fpm (0.4 – 50m/s): 212 – 392ºF
(100 – 200C) |
|
-- |
138
– 9840fpm (0.7 – 50m/s): 392 – 572ºF
(200 – 300ºC) |
|
-- |
197
– 9840fpm (1.0 – 50m/s): 572 – 752ºF
(300 – 400ºC) |
|
Temp. Compensation
Accuracy (Air Velocity) |
Less than 984fpm (5m/s): +/-10%F.S.
984fpm(5m/s) to 9,840fpm(50m/s): +/-6%F.S. |
Less than 984fpm (5m/s): +/-15%F.S.
984fpm(5m/s) to 9,840fpm(50m/s): +/-10%F.S. |
|
Heat-resisting of Cable |
Teflon coating (Probe side): 392ºF
(200ºC)
Vinyl code (Extension cable): 176ºF
(80ºC) |
|
Length of Cable |
Teflon coating 4.9ft (1.5m)
Vinyl code 16.4ft (5m) |
Teflon coating 7.5ft (2.3m)
Vinyl code 32.8 (10m) |
|
Extension Rod (Option) |
0.65”(MAX)x31.5” (16.5x800mm) |
0.87”(MAX)x81.5” (22x2070mm) |
|
Probe Dimension |
Dimension:
φ0.43”x8.2”
(11x208mm) |
Dimension:
φ0.43”x39.4”
(11x1000mm) |
|
Display |
Digital (simultaneous display of air velocity and temperature) |
|
Input/output Terminal |
Remote terminal: START/STOP key |
|
Analog output terminal: Output voltage 0 – 1 V, Output
impedance 47 Ω |
|
Digital output terminal: RS-232C (serial interface) |
|
Power Supply |
Dry
battery drive: US type (1.5V x 6 pcs = 9V), Alkaline battery, Mn
battery
AC
adapter: 12.5V, 450mA (AC100V +/- 10%, 50/60Hz) |
|
Operating Temp. |
41
– 104ºF
(5 – 40ºC) |
|
Battery Life |
Approx. 8hours |
|
Dimensions |
8.7” x 3.3” x 5.9” 220 x 85 x 150 mm |
|
Weight |
Main body: Approx. 4.0lbs (1.8kg), Probe Model 0203:
7.1oz(200g), Model 0204: 17.6oz (500g) |